CyberScan
pH/Ion/Conductivity Meters
 |
Eutech's
bench meters are equipped with Application
Specific Integrated Circuit (ASIC) which allows
user-friendly interface. The economy meters are
no-frills meters simple to use with user-friendly
features.
The advanced meters offer advanced features of
high performance meters without the associated
complexity. The microprocessor-based meters
allows user customization through SETUP function,
ensuring easy and simple operation. |
Comprehensive
program options such as real-time clock, two-way
communication link, timer, alarm and non-volatile memory
backup function are built-in standards. Besides pH
measurement, you can measure ORP/Redox and Ion
Concentration (direct reading with pH 2500) with
comprehensive range of ORP/Ion Selective Electrodes (ISE).
The
pH/Ion510 has 50 memory data sets, up to 3-Point
calibration with choice of USA or NIST with auto-buffer
recognition minimizing calibration errors.
Auto-ranging
CON500 meter allows customizable functions that can
suit individual needs for routine laboratory
measurement.
Features
User
customization
is easy with comprehensive options to suit your specific
needs for enhanced flexibility and convenience. Some
customizable functions includes selectable pH
resolution, clear/reset calibration data and memory, set
date/time, ready, data log/print criteria, communication
protocol, etc.
Multi-point
Push-button Calibration with
up to 5 points for full range accuracy, with auto buffer
recognition minimizes calibration errors and built-in
ATC, ensures reliability of measurement. A special CAL
EDIT function allows non-standard pH buffer values
to be input for wider calibration options.
Non-volatile
Memory for 16 Data Sets can
be stored and recalled in a Last-In-First-Out (LIFO)
sequence. Sensor calibration values and stored data is
retained, even when meter is powered-off or accidentally
disconnected.
Direct Ion
Concentration Measurement is
possible with CyberScan pH2500, ION 510 using appropriate half
cell or combination ion selective electrodes (ISE).
Meter displays slope of ISE, and allows minimum 2-point
and maximum 5-point calibration for unprecedented
full-range accuracy.
Durable
Electrode with Built-in ATC utilizes
2 platinum rings epoxy-body
Conductivity electrode (cell constant of 1.0) specially
designed for low volume measurement with minimum air
bubble entrapment. It has excellent chemical durability
and ideal for most applications. (CON 500)
Specifications
of CyberScan pH 1000 & 2500 Meters
|
Product
Code
|
EC-PH1000 |
EC-PH2500 |
| pH
Range |
-2.00
to 16.00 pH |
| Resolution
& Accuracy |
0.01
& ±0.01 pH |
| mV
Range |
±399.9
mV (extended); ±1999 mV |
| Resolution
& Accuracy |
0.1
mV & ±0.2 mV (within ±399.9 mV);
1 mV & ±2 mV (beyond ±400 mV) |
| Ion
Concentration Range |
Not
Available |
0.00
to 9999 |
| Resolution
& Accuracy |
- |
±1
least significant digit |
| Temperature
Range |
0.0
to 100.0 °C |
| Resolution
& Accuracy |
0.1
°C & ±0.5 °C |
| No.
of pH Calibration Points |
Up
to 5 points (Standard & Editable Buffers) |
| pH
Buffer Options |
pH
1.68, 4.01, 7.00, 10.01, 12.45 |
| No.
of Ion Calibration Points |
Max.
5, min 2. points |
| Temperature
Compensation |
Automatic
/ Manual (0 to 100 °C) |
| pH
Slope & Offset Display |
Yes |
| Memory |
16
Sets |
| Print
Output |
Recorder,
RS232C |
| Special
Features |
Timer;
Alarm (High & Low); Automatic
Data-logging/Printing |
| Additional
Functions |
Averaging/Stability;
Self-diagnostic;
Hold & Auto-Power Off |
| Power
Requirements |
9V
AC/DC Adapter |
| PRICE
(RS) |
POR |
Specifications
of CyberScan pH/Ion 510 Meters
|
Product
Code
|
EC-PH510/510S |
EC-ION510/510S |
| pH
Range |
0.00
to 14.00 pH |
| Resolution
& Accuracy |
0.01
& ±0.01 pH |
| Ion
Concentration Range |
Not
Available |
0.01
to 1999 ppm |
| Resolution
& Accuracy |
- |
0.01
ppm (0.01 to 0.99 ppm);
0.1 ppm (1.0 to 199.9 ppm);
1 ppm (200 to 1999 ppm)
& ± 1% Full Scale |
| mV
Range |
±199.9
mV; ±1999 mV |
| Resolution
& Accuracy |
0.1
mV & ±0.2 mV (within ±199.9 mV);
1 mV & ±2 mV (beyond ±200 mV) |
| Temperature
Range |
0.0
to 100.0 °C |
| Resolution
& Accuracy |
0.1
°C & ±0.5 °C |
| No.
of calibration points |
Up
to 3 points |
Up
to 3 points (pH & Ion) |
| pH
buffer standards/options |
USA
(pH 4.01, 7.00, 10.01) or
NIST (pH 4.01, 6.86, 9.18) |
| Temperature
Compensation |
Automatic
/ Manual (0 to 100 °C) |
| Memory |
50
Data Sets |
| Recorder
Output |
Yes |
| Special
Functions |
Averaging/Stability;
Self-diagnostic;
Hold & pH Slope/Offset Display |
| Power
Requirements |
9V
AC/DC Adapter |
| PRICE
(RS) |
POR |
Specifications
of CyberScan CON500 Meter
| Product
Code |
EC-CON500 |
| Conductivity
Range |
0
to 19.99, 199.9, 1999 µS/cm;
19.99, 199.9 mS/cm |
| Resolution |
0.05%
of Full Scale |
| Accuracy |
±
1% Full Scale or ± 1 digit |
| Temperature
Range |
0.0
to 100.0 °C |
| Resolution
& Accuracy |
0.1
°C & ± 0.5 °C |
| Normalization
Temperature |
20.0
or 25.0 °C (selectable) |
| Cell
Constant |
0.1,
1.0 or 10.0 (selectable) |
| Temperature
Coefficient |
0.0
to 10.0 %/°C (adjustable) |
| No.
of Calibration Points |
5
(1 per range) |
| Memory |
16
Data Sets |
| Special
Functions |
Auto-ranging;
Averaging/Stability; Self-diagnostic; Hold
& Auto-Power Off |
| Power
Requirements |
9V
AC/DC Adapter |
| PRICE
(RS) |
POR |
|