CYBERSCAN TURBIDITYMETER TB 1000 SERIES INFRAED AND WHITE LIGHT
 |
Using infrared and tungsten white light sources, Eutech's turbidity meters deliver +2% full scale accuracy and a resolution of 0.01
NTU.
These meters are excellent analytical tools for research, quality control and routine measurements.
|
AUTO-RANGING UP TO 1000 NTU
The meter senses turbidity level of measured sample and automatically adjusts to the appropriate measurement range, eliminating guesswork.
PUSH-BUTTON CALIBRATION WITH AUTO ALERT CALIBRATION
PROMPT
Simple calibration procedure is initiated at the touch of buttons and the instrument automatically prompts the user when next calibration is required.
RS232 OUTPUT
Uni-directional RS232 communication interface allows NTU readings (with time and date) to be downloaded to a serial printer or data recorder.
POUR-THROUGH ASSEMBLY
This optional assembly module allows instantaneous sampling through the measurement chamber without having to transfer samples into cuvettes saving time.
TWO MODELS :
EC-TBDW1000 (TURBIDITY METER WITH WHITE LIGHT SOURCE)
EC-TBDIR1000 (TURBIDITY METER WITH INFRARED LIGHT SOURCE)
APPLICATIONS
Drinking * Waste water treatment * Petrochemical * Electroplating * Beverage production
Specifications
CyberScan TB1000 Series
| Principle of Operation |
Nephelometric non-ratio |
| Range |
0 to 1000 NTU ( Auto-ranging) |
| Resolution |
0.01 NTU ( 0 to 9.99 NTU)
0.1 NTU (10 to 99.9 NTU)
1 NTU (100 to 1000 NTU) |
| Accuracy |
+2% of reading +0.01 NTU |
| Repeatability |
+1% of reading +0.01 NTU |
| Calibration |
1 to 3 point (automatic) |
| Response Time |
<6 seconds |
| Sample Volume |
30 ml (27 ml minimum) |
| Light Source |
White Light (Tungsten) or Infrared |
| Real-time clock |
Date and time |
| Output |
RS 232 |
| Operating Temperature |
0 to 40 deg C |
| Dimensions (LxBxH) |
27x25x10 cm |
| Weight |
25 kg |
| Power requirements |
12VDC power adapter, 500mA |
| PRICE
(RS) |
POR |
|